PRODUCTS

Coating Thickness Gauge

Serial number
CG510PL.
没有此类产品
immediate inquiry
Product description
Specifications

Coating Thickness Gauge

Features:

●Two thickness measuring methods adopted, the gauge can be applied to thickness measuring of non-magnetic coating on magnetic metal substrate as well as non-conductive  coating on non-magnetic metal substrate;

●5 types of probes can be used (F400, F1, F10, N1, CN02,);

●2 measuring modes are available: continuing measuring mode(CONTINUE)and single

   measuring mode(SINGLE);

●2 operation modes are available: direct mode (DIRECT) and batch mode (A-B);

●5 statistic values:mean value(MEAN), max. value(MAX), min. value(MIN), numbers

   of measuring(NO.), standard deviation(S.DEV);

●2 methods can be used to calibrate the gauge, and the system error of the probe can be corrected

  by use of basic calibrating method:

●Storage function: 495 measuring values can be stored;

●Deletion function: delete the single questionable data occurring in measuring, as well as all of the

  data in memory area to perform the new measuring;

●Limit can be set: capable of alarming automatically for measuring values out of limit; and a batch

  of measuring values can be analyzed via histogram;

●Printing function: measuring value, statistic value, limit and histogram can be printed;

●Capable of communication with PC: measuring values and statistic values can be sent to PC to

   perform the further process for data;

●Indication of low voltage;

●Buzzing indication in the course of operation;

●With the function of error warning, error warning can be carried out through display or buzzing;

●Two shut down modes are available: manual shut down mode and automatic shut down mode;

INTERNATIONAL STANDARDS:

  ASTM B 499, ASTM D 1400, BS 3900-C5-6B, DIN 50981,DIN 50984, ISO 1461,ISO 2360

Corresponding parameter set not found, please add it in property template of background

 Technical parameters

Probe model

F400

F1

F1/90°

F5

F10

Operating principle

Magnetic induction

Measuring range (mm)

0~400

0~1250

0~5000

0~10000

Low range resolution(mm)

0.1

0.1

1

10

Accuracy

One-point calibration (mm)

±(3%H+1)

±(3%H+5)

±(3%H+10)

Two-point calibration (mm)

±((1~3)%H+0.7)

±((1~3)%H+1)

±((1~3)%H+5)

±((1~3)%H+10)

Measuring conditions

Min curvature radius (mm)

Convex

1

1.5

Flatten

5

10

Diameter of the min area (mm)

F3

F7

F7

F20

F40

Critical thickness of substrate (mm)

0.2

0.5

0.5

1

2

 

Probe model

N400

N1

N1/90°

CN02

N10

Operating principle

Eddy current

Measuring range (mm)

0~400

(Chrome on copper

 

0~1250

0~40)

10~200

0~10000

Low range resolution(mm)

0.1

0.1

1

10

Accuracy

One-point calibration (mm)

±(3%H+0.7)

±(3%H+1.5)

±(3%H+1)

±(3%H+25)

Two-point calibration (mm)

±[(1~3)%H+0.7]

±[(1~3)%H+1.5]

___

±[(1~3)%H+25]

Measuring conditions

Min curvature radius (mm)

Convex

1.5

3

Flatten

Only flatten

25

Diameter of the min area (mm)

F4

F5

F5

F7

F50

Critical thickness of substrate (mm)

0.3

0.3

0.3

No limit

50mm aluminum foil

    Note: H——Nominal value

    Table 1 for probe selection reference

 

Coating

Substrate

 

Non-magnetic coating of organic material (such as painting, finishes, enamel, porcelain enamel, plastic, anodization and etc.

Coating thickness≤100mm

Coating thickness > 100mm

Magnetic metal such as iron, steel and etc.

 

Measured area

Dia. > 30mm

 

Probe F1     0~1250mm

Probe F400   0~400mm

Probe F1     0~1250mm

Probe F5     0~5mm

Probe F10    0~10mm

Measured area

Dia. < 30mm

Probe F400   0~400mm

Probe F1     0~1250mm

Probe F400   0~400mm

Non-ferrous metal such as copper, aluminum, brass, zinc, tin and etc.

Measured area

Dia.> 10mm

Probe N1     0~1250mm

Probe N400   0~400mm

 

Probe N1     0~1250mm

Probe N10    0~10mm

Probe N400   0~400mm

 

Measured area

Dia. < 10mm

Probe N1     0~1250mm

Probe N400   0~400mm

Probe N1     0~1250mm

Probe N400   0~400mm

    Table 2 for probe selection reference

 

Coating

Substrate

 

Non-magnetic non-ferrous metal coating (such as chrome, zinc, aluminum, copper, tin, silver and etc.)

Coating thickness≤ 100mm

Coating thickness > 100mm

 

Magnetic metal such as iron, steel and etc.

Measured area

Dia. > 30mm

Probe F1     0~1250mm

Probe F400   0~400mm

Probe F1     0~1250mm

Probe F5     0~5mm

Probe F10    0~10mm

Probe F400   0~400mm

Measured area

Dia. < 30mm

Probe F1     0~1250mm

Probe F400   0~400mm

 

Probe F1    0~1250mm Probe F400   0~400µm

Non-ferrous metal such as copper, aluminum, brass, zinc, tin and etc.

Measured area

Dia. > 10mm

Only for chrome coating on copper

Probe N1     0~1250mm

Probe N400   0~40mm

--------

Measured area

Dia. < 10mm

--------

--------

Non-metal such as plastic, printing circuit and etc.

Measured area

is big

Probe CN02   10~200mm

Probe CN02   10~200mm

 